NuGenesis Archive Agent 로그 파일에 "Exception of type 'System.OutOfMemoryException' was thrown" 오류가 표시됨 - WKB59028

Article number: 59028To English version

증상

  • Empower 데이터가 NuGenesis에 아카이빙되지 않습니다. 8 SR-2

환경

  • NuGenesis 8 SDMS
  • NuGenesis 8 Data Adapters Release 3 HF1
  • Empower 3

원인

JCAMP 변환이 활성화된 Empower 프로젝트를 아카이브하려고 시도했습니다. 대상 프로젝트가 너무 커서 변환 모듈이 처리할 수 없습니다.

수정 또는 해결 방법

이 문제를 해결하려면 Empower Remote DCOM 서버에 DAR3-HF2를 설치하십시오.

추가 정보

Archive Agent의 로그 파일인 "Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log"에 다음 메시지가 나타납니다.

Message........: 2018-01-16 07:22:21: Beginning execution of archive request 1.
Message........: 2018-01-16 07:22:21: Server: NUG, Project: Empower, Template: EmpClient.
Message........: 2018-01-16 07:22:22: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 1/16/2018 7:22:21 AM)(End Time = 1/16/2018 7:22:22 AM)(Total Time = 0h 0m 1s)}.
"E ","2018/01/16 08:01:04 AM ","-2147417851 ","-2147417851 ","The server threw an exception. (0x80010105 (RPC_E_SERVERFAULT)): ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","NGPerformConvTasks ","1 ",""
"E ","2018/01/16 08:01:18 AM ","7 ","7 ","Exception of type 'System.OutOfMemoryException' was thrown.:EMP3.CUSTOMERS\GSK\EVR_MV_SA_MDI_FP_FAST_08_03 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","ExtractKeys ","1 ",""
"E ","2018/01/16 08:01:18 AM ","0 ","0 ","Error extracting keys for the project:EMP3.CUSTOMERS\GSK\EVR_MV_SA_MDI_FP_FAST_08_03 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/01/16 08:01:18 AM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/01/16 08:01:18 AM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-CL ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-01-16 08:01:19: EMC: PerformPreProcess error.
Message........: 2018-01-16 08:01:19: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 1/16/2018 7:22:22 AM)(End Time = 1/16/2018 8:01:19 AM)(Total Time = 0h 38m 57s)}.
Message........: 2018-01-16 08:01:19: Next scheduled runtime for archive request 1 is 2018-01-16 11:00:42.
 

id59028, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG