NuGenesis Archive Agent 로그 파일에 "The server threw an exception" 오류가 표시됨 - WKB59026
증상
- Empower 데이터가 NuGenesis 8 SR2에 아카이빙되지 않습니다.
환경
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters Release 3 HF1
- Empower 3
원인
JCAMP 변환이 활성화된 Empower 프로젝트를 아카이브하려고 시도했습니다. 대상 프로젝트가 너무 커서 변환 모듈이 처리할 수 없습니다.
수정 또는 해결 방법
Empower Remote DCOM 서버에 DAR3-HF2를 설치합니다.
추가 정보
Archive Agent의 로그 파일인 "Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log"에 다음 메시지가 나타납니다.
Message........: 2018-01-16 11:41:12: Beginning execution of archive request 1.
Message........: 2018-01-16 11:41:12: Server: NUG, Project: Empower, Template: EmpServer.
Message........: 2018-01-16 11:41:13: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 1/16/2018 11:41:12 AM)(End Time = 1/16/2018 11:41:13 AM)(Total Time = 0h 0m 1s)}.
"E ","2018/01/16 11:45:55 AM ","-2147417851 ","-2147417851 ","The server threw an exception. (0x80010105 (RPC_E_SERVERFAULT)): ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","NGPerformConvTasks ","1 ",""
Message........: 2018-01-16 11:45:55: EMC: PerformPreProcess error.
"E ","2018/01/16 11:45:55 AM ","5 ","5 ","The socket connection was aborted. This could be caused by an error processing your message or a receive timeout being exceeded by the remote host, or an underlying network resource issue. Local socket timeout was '7.23:00:00'. While extracting keys for file : \\WB-EMP-SRV\SDMSarchive\EMP3\TEST\CONVERSION\01 EVR_MV_SA_MDI_FP_FAST_08_03\01 EVR_MV_SA_MDI_FP_FAST_08_03.EXP ","Waters.SDMS.EmpowerDAProxy ","ClsEmp2FAExtract ","ExtractKeys ","1 ",""
"E ","2018/01/16 11:45:55 AM ","5 ","5 ","Invalid procedure call or argument While extracting keys for file : \\WB-EMP-SRV\SDMSarchive\EMP3\TEST\CONVERSION\01 EVR_MV_SA_MDI_FP_FAST_08_03\01 EVR_MV_SA_MDI_FP_FAST_08_03.EXP ","NG80EVSEmp2PROXY ","clsEmp2FAExtract ","ExtractKeys ","1 ",""
Message........: 2018-01-16 11:45:55: ExecuteTemplateCommand. Unable to extract keys using objiCommonKEE.iExtractKeys.
Message........: 2018-01-16 11:46:04: FillTransactionDO. (5): Invalid procedure call or argument.
Message........: 2018-01-16 11:46:04: EMC: FillTransactionDO failure.
"M ","2018/01/16 11:46:08 AM ","0 ","0 ","Archived Project:[EMP3.TEST\CONVERSION\01 EVR_MV_SA_MDI_FP_FAST_08_03] "," "," "," ","0 ",""
Message........: 2018-01-16 11:46:08: ExecuteTemplateCommand: The group is rolled back because of errors.
Message........: 2018-01-16 11:46:08: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 1/16/2018 11:41:13 AM)(End Time = 1/16/2018 11:46:08 AM)(Total Time = 0h 4m 55s)}.
Message........: 2018-01-16 11:46:08: Next scheduled runtime for archive request 1 is 2018-01-17 09:00:56.
id59026, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG