ACQUITY UPLC H-Class 에서 머무름 시간 변동과 넓은 피크가 관찰됨 - WKB61930

Article number: 61930To English version

증상

  • 머무름 시간 변동
  • 압력 리플이 불규칙함

환경

  • ACQUITY APC
  • ACQUITY Arc
  • ACQUITY Arc Premier
  • ACQUITY UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class Quaternary Solvent Manager(QSM)
  • ACQUITY UPLC H-Class Bio Quaternary Solvent Manager(QSM)
  • ACQUITY UPLC H-Class Plus QSM(Quaternary Solvent Manager)
  • ACQUITY UPLC H-Class PLUS Bio Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Premier 시스템

원인

용매 용기 필터가 막혔습니다.

수정 또는 해결 방법

용매 용기 필터(700003616)를 교체하십시오.

추가 정보

id61930, A-30SDM, MLYNX, MLYNXV41, retention time, SUPMM, UPAQSM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARPFTN, UPARPQSM, UPARPTUV, UPARQSM, UPARSMFTN, UPBINARY, UPBINARYUP, UPBSM+, UPBSM+IVD, UPBSMBIO+, UPIBSM, UPIBSM+, UPPBSM, UPPDAARCB, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO