HDI 1.4에 표시되는 MaldiChrom 프로세싱 오류 - WKB46116
증상
- HD Imaging (HDI) v1.4에서 데이터를 프로세싱할 때 오류가 관찰됩니다
- 오류 예시:
MaldiChrom v2.14 Feb-2016
Copyright 2008-2016 Micromass UK Ltd.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Tolerance' in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Flatten Edges’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445)] Failed to find key 'Type’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
[PROCPARS:gpps.cpp(435)] Failed to find section [TOFACM MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini. [PROCPARS:gpps.cpp(445]] Failed to find key ’Centroiding Type’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process \MaldichromParameter.ini.
[PROCPARS:gpps.cpp(445]] Failed to find key ’Use Areas’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.
0 lockmass adjustments to be applied
MSE Level 0
환경
- HDI v1.4 소프트웨어
- SYNAPT G2-S/Si MALDI
원인
이러한 메시지는 정상이며 .ini 파일에 매개 변수가 지정되지 않았기 때문에 나타납니다. 그 후 기본 매개 변수가 사용됩니다.
수정 또는 해결 방법
무시해도 됩니다. 메시지는 정상입니다.
추가 정보
id46116, MALDIG2SI, MALDIG2SI, MALDIMICRO, SYNG2SIHD