NuGenesis Archive Agent 로그 파일에 "PerformProjBackup failed" 오류가 표시됨 - WKB59046
증상
- Empower 데이터가 NuGenesis에 아카이빙되지 않습니다.
환경
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
원인
Archive Template Request Line에 지정된 대로 Empower 사용자에게 "Alter Any Project" 또는 "Backup Project" 권한이 없습니다.
수정 또는 해결 방법
Empower에서 사용자에게 적절한 권한을 할당합니다.
추가 정보
Archive Agent의 로그 파일인 "Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log"에 다음 메시지가 나타납니다.
Message........: 2018-12-18 14:05:43: Beginning execution of archive request 1.
Message........: 2018-12-18 14:05:43: Server: NUG, Project: Empower, Template: EmpServer.
Message........: 2018-12-18 14:05:44: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 12/18/2018 2:05:43 PM)(End Time = 12/18/2018 2:05:44 PM)(Total Time = 0h 0m 1s)}.
"E ","2018/12/18 02:05:47 PM ","-2147220896 ","-2147220896 ","-2147220896Exception from HRESULT: 0x80040260 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","BackupProjectTK ","1 ",""
"E ","2018/12/18 02:05:47 PM ","-1 ","-1 ","PerformProjBackup failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:05:47 PM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:05:47 PM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-SRV ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-12-18 14:05:47: EMC: PerformPreProcess error.
Message........: 2018-12-18 14:05:50: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 12/18/2018 2:05:47 PM)(End Time = 12/18/2018 2:05:50 PM)(Total Time = 0h 0m 3s)}.
Message........: 2018-12-18 14:05:50: Next scheduled runtime for archive request 1 is 2018-12-18 14:06:32.