NuGenesis Archive Agent 로그 파일에 "Project Locking failed" 오류가 표시됨 - WKB59051
증상
- Empower 데이터가 NuGenesis에 아카이빙되지 않습니다.
환경
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
원인
File Capture Template의 Request Line에 지정된 Empower 사용자에게는 Empower "Lock Project" 권한이 없습니다.
수정 또는 해결 방법
Empower에서 사용자에게 적절한 권한을 할당합니다.
추가 정보
Archive Agent의 로그 파일인 "Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log"에 다음 메시지가 나타납니다.
Message........: 2018-12-18 14:31:44: Beginning execution of archive request 1.
Message........: 2018-12-18 14:31:44: Server: NUG, Project: Empower, Template: EmpServer.
Message........: 2018-12-18 14:31:45: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 12/18/2018 2:31:44 PM)(End Time = 12/18/2018 2:31:45 PM)(Total Time = 0h 0m 1s)}.
"E ","2018/12/18 02:31:52 PM ","-2147220896 ","-2147220896 ","Exception from HRESULT: 0x80040260 ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","LockProject ","1 ",""
"E ","2018/12/18 02:31:52 PM ","-1 ","-1 ","Project Locking failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:31:52 PM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/12/18 02:31:52 PM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-SRV ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-12-18 14:31:52: EMC: PerformPreProcess error.
"M ","2018/12/18 02:31:54 PM ","0 ","0 ","Archived Project:[EMP3.MANY\Tea Group\Tea 002] "," "," "," ","0 ",""
Message........: 2018-12-18 14:31:54: Archive scan summary for command id 1: {"NUG","Empower","EmpServer"} {"Waters Corp.","Empower-3 Projects","0"} {Summary : (Command ScanPath = \\WB-EMP-SRV\SDMSarchive)(Total Scanned Files = 14)(Total Archived Files = 14)(Total Bytes = 3570891)(Start Time = 12/18/2018 2:31:45 PM)(End Time = 12/18/2018 2:31:54 PM)(Total Time = 0h 0m 9s)(Errors = 0)}.
Message........: 2018-12-18 14:32:03: Next scheduled runtime for archive request 1 is 2018-12-18 14:32:32.
id59051, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG, 프로젝트