NuGenesis LMS 섹션 템플릿에 파일을 추가할 때 "Not enough quota is available to process this command" 오류가 표시됨 - WKB13915
증상
- NuGenesis LMS Section Management의 섹션 템플릿에 파일을 추가하려고 시도할 때 다음 오류가 나타납니다.
Not enough quota is available to process this command
at MS.Win32.UnsafeNativeMethods.PostMessage(HandleRef hwnd, WindowMessage msg, IntPtr wparam, IntPtr lparam)
at System.Windows.Interop.HwndTarget.UpdateWindowSettings(Boolean enableRenderTarget, Nullable`1 channelSet)
at System.Windows.Interop.HwndTarget.UpdateWindowPos(IntPtr lParam)
at System.Windows.Interop.HwndTarget.HandleMessage(WindowMessage msg, IntPtr wparam, IntPtr lparam)
at System.Windows.Interop.HwndSource.HwndTargetFilterMessage(IntPtr hwnd, Int32 msg, IntPtr wParam, IntPtr lParam, Boolean& handled)
at MS.Win32.HwndWrapper.WndProc(IntPtr hwnd, Int32 msg, IntPtr wParam, IntPtr lParam, Boolean& handled)
at MS.Win32.HwndSubclass.DispatcherCallbackOperation(Object o)
at System.Windows.Threading.ExceptionWrapper.InternalRealCall(Delegate callback, Object args, Int32 numArgs)
at MS.Internal.Threading.ExceptionFilterHelper.TryCatchWhen(Object source, Delegate method, Object args, Int32 numArgs, Delegate catchHandler)
환경
- NuGenesis 8 LMS SR2
- NuGenesis 8 LMS FR1
- NuGenesis 8 ELN SR1
원인
NuGenesis LMS 클라이언트 시스템의 페이지 파일이 너무 작습니다.
수정 또는 해결 방법
- Windows Control Panel > System을 엽니다.
- Advanced 탭을 선택하고 Performance 아래에서 Settings를 클릭합니다.
- Performance Options 창에서 Advanced 탭을 선택하고 Virtual memory에서 Change를 클릭합니다.
- "Automatically manage paging file size for all drives" 옵션을 선택합니다.
- 모든 대화 상자에서 OK를 클릭하고 요청하면 시스템을 다시 시작합니다.
추가 정보
위 단계는 최신 버전의 Windows에서 변경될 수 있습니다. 사용자가 Citrix를 통해 LMS를 실행하는 경우 Citrix 서버에서 해당 단계를 실행합니다.