Waters 시스템의 열부하는 얼마입니까? - WKB45002
Article number: 45002To English version
환경
- ACQUITY APC
- ACQUITY APC Isocratic Solvent Manager
- ACQUITY APC Sample Manager - FTN
- ACQUITY Arc
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC
- ACQUITY UPLC BSM
- ACQUITY UPLC Column Manager
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class Quaternary Solvent Manager(QSM)
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- PATROL UPLC 프로세스 분석 시스템
- ACQUITY UPLC 컬럼 히터
- ACQUITY UPLC PDA 검출기
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS(SM-FTN PLUS)
- ACQUITY UPLC Sample Organizer
- ACQUITY UPLC Tunable UV 검출기
- ACQUITY UPLC FLR 검출기
- ACQUITY UPLC Online SPE Man
- 3100 질량검출기
- SQ 검출기
- SQ 검출기 2
- TQ 검출기
- ACQUITY TQD
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- BioAccord 시스템
- ACQUITY RDa 검출기
- SYNAPT G2-S MS/HDMS
- SYNAPT G2-Si MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- Vion IMS QTof
- Xevo QTof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
답변
Site Preparation Guide("설치 준비 안내서")는 각 모듈의 전력 소비량에 대한 정보를 제공합니다.
Site Preparation Guide("설치 준비 안내서")의 Power Requirements(전원 요구사항) 섹션까지 아래로 스크롤합니다.
요약 표에는 "Typical Power Consumption"(일반 전력 소비량)이 나와 있습니다. 단위는 W(와트) 또는 VA(볼트 - 암페어)중 하나일 수 있습니다. W = VA(역률이 1이기 때문에)
W 또는 VA에서 Btu/hr로 전환합니다. W * 3.413 = Btu/hr
냉장(냉각) 1톤 = 12000Btu/시간
총 냉각 요건을 결정하려면 실험실에 있는 각 시스템의 열부하를 계산해야 합니다.
추가 정보
귀하에게 mplementation Manager가 할당되었다면 해당 시스템의 Site Preparation Guide("설치 준비 안내서")를 제공받을 수 있습니다. 아직 시스템을 구매하지 않았거나 Implementation Manager를 할당받지 못한 경우, MS_Service@waters.com에 문의하여 원하는 모듈의 안내서를 요청하십시오.
id45002, column heater