어떤 샘플 콘이 어떤 MS와 함께 사용됩니까? - WKB92340
Article number: 92340To English version
환경
- 3100 질량검출기
- SQ 검출기
- SQ 검출기 2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQ Absolute
- Xevo TQ-S micro
- Xevo TQ-S cronos
- Xevo QToF
- Xevo G2 ToF
- Xevo G2 QToF
- Xevo G2-S
- Xevo G2-XS
- SYNAPT HDMS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Vion IMS QTof
- SELECT SERIES Cyclic IMS
답변
| 콘 직경(mm) | 엔지니어링 파트 넘버 | Waters 파트 넘버 | 사용 기기: |
0.25 |
405026128 | 700012275 | Xevo TQ-S cronos |
| 0.35 | 405006555 | 700004219 | Xevo Tof, Xevo QToF, Xevo G2 ToF, Xevo G2 QToF, Xevo TQ-MS, SYNAPT G2 |
| 0.36 | M946699BD1 | M946699BD1-S | SYNAPT HDMS |
| 0.45 | 289009523 405017868 |
700010289 | Xevo TQ-S micro |
| 0.50 | 405005883 | 700003055 | 3100 질량검출기, SQ 검출기, TQD |
| 0.50 | 405011167 | 700005997 | Xevo TQD. SQD2 |
| 0.67 | 405013693 | 700009087 | Xevo G2-S QToF, Xevo G2-XS QToF, Vion |
| 0.80 | 405010077 | 700005350 | Xevo TQ-S, SYNAPT G2-S, SYNAPT G2-Si |
| 0.90 | 289010932 405018793 |
700011217 | Xevo TQ-XS, SYNAPT XS, CycIMS, TQ Absolute |
추가 정보
이는 Waters 질량 분석 기기와 관련된 콘의 크기가 다름을 알리기 위한 것입니다. 이들은 기기 간에 상호 교환할 수 없습니다. 그 용도에 맞는 기기에 적절한 콘만 사용해야 합니다. 첨부된 사진을 참조하여 파트 넘버의 위치를 확인하십시오.


id92340, A-SQ, A-SQ, CYCIMS, MALDIG2SI, sample cone, SQD, SQD2, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD