BSM 변환기의 Outlet 피팅(Fitting)이 부러졌습니다. - WKB262365
Article number: 262365To English version
증상
- 압력 변동
- Dynamic leak 테스트 실패
환경
- ACQUITY UPLC
- I-Class BSM
- H-Class QSM
- H-Class Bio QSM
- ACQUITY Premier BSM
- nanoACQUITY
- M-Class
- ACQUITY ISM
- UPC2
원인
BSM 변환기의 Outlet 피팅(Fitting)이 부러졌습니다.
수정 또는 해결 방법
1. 손상된 변환기를 교체합니다.
- 700006045 - Transducer, 20K PSI for I-Class systems
- 700002594 - Pressure Transducer Assembly, 15K psi (ACQUITY UPLC, H-Class, H-Class Bio, nanoACQUITY, UPC2 (B side), ACQUITY Premier)
- 700006045 - Transducer Assembly ISM, M-Class
- 700009006 - Transducer Assembly, Pump A, 15K psi, Parylene-Coated, UPC2 (A side)
2. 손상된 튜빙을 교체합니다.
Aux용 펌프 B Accumulator(파트 넘버 700009849)의 Outlet 튜빙 손상되었습니다.
추가 정보
id262365, DLT, leak test, UPAISM, UPBSMIVD, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPIBSM, UPIBSM+, UPIBSMIVD, UPISM, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPLCCOLHTR, UPMASM, UPMASM, UPMBSM, UPMSMFL, UPNAYXPMP, UPNBINARY, UPNSAMMGR, UPNSOLMGR, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR, 누수 테스트