"Loader.exe"를 통해 수동으로 펌웨어를 로드하는 방법 - WKB259491

Article number: 259491To English version

목적 또는 목표

"Loader.exe"를 통해 수동으로 펌웨어를 로드합니다.

환경

  • Empower 3
  • MassLynx
  • ACQUITY APC
  • ACQUITY APC Isocratic Solvent Manager
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY Arc
  • ACQUITY Arc Premier
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC Column Manager
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class Quaternary Solvent Manager(QSM)
  • ACQUITY UPLC H-Class Bio Quaternary Solvent Manager(QSM)
  • ACQUITY UPLC H-Class Plus QSM(Quaternary Solvent Manager)
  • ACQUITY UPLC H-Class PLUS Bio Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC M-Class
  • nanoACQUITY UPLC
  • PATROL UPLC 프로세스 분석 시스템
  • ACQUITY UPLC Process Sample Manager
  • ACQUITY UPLC PDA 검출기
  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • ACQUITY UPLC Sample Organizer
  • ACQUITY UPLC Tunable UV 검출기
  • ACQUITY UPLC FLR 검출기
  • ACQUITY UPLC RI 검출기
  • ACQUITY Premier 시스템
  • 2414 RI 검출기
  • 2424 ELS 검출기
  • 2475 Multi-Wavelength FLR 검출기
  • 2489 UV/Vis 검출기
  • 2998 PDA 검출기
  • ACQUITY UPLC Tunable UV 검출기
  • ACQUITY UPLC PDA 검출기
  • ACQUITY UPLC PDA eλ 검출기
  • ACQUITY UPLC FLR 검출기
  • ACQUITY UPLC ELS 검출기

절차

  1. 펌웨어 폴더에 있는 "loader.exe" 파일을 찾아 엽니다.
    • MassLynx: C:\Program Files (x86)\ Waters Instruments\Firmware
    • Empower: C:\ Empower\ Waters Instruments\Firmware

Loader.JPG

2. "Scan for instruments" 확인란을 선택합니다.

3. 펌웨어를 수동으로 변경할 기기(또는 공백인 경우 CPU2000)를 찾습니다.

4. 로드하려는 특정 .sft 파일을 찾습니다.

5. Load를 선택하고 완료될 때까지 기다립니다. 지시가 있을 때까지 모듈의 전원을 껐다가 켜지 마십시오.

추가 정보

id259491, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, MLYNX, MLYNXV41, SUP, SUPMM