ACQUITY UPLC 시스템의 고스트 피크(Ghost peak) - WKB200730
Article number: 200730To English version
증상
- Blank injection의 고스트 피크
환경
- ACQUITY Arc
- ACQUITY Arc Bio
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
원인
오염된 인라인(In-line) 필터 카트리지
수정 또는 해결 방법
인라인(In-line) 필터 카트리지를 교체합니다.
추가 정보
주문 정보:
- 700002913, ACQUITY Arc
- 700011679, ACQUITY Arc Bio
- 700002913, ACQUITY UPLC
- 700005173, ACQUITY UPLC H-Class
- 700005259, ACQUITY UPLC H-Class Bio
- 700002913, ACQUITY UPLC I-Class
- 700005173, ACQUITY UPLC H-Class PLUS
- 700005259, ACQUITY UPLC H-Class PLUS Bio
- 700002913, ACQUITY UPLC I-Class PLUS
id200730, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPDAARCB, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMPMGR