ACQUITY 30cm 컬럼 히터/쿨러가 작동하지 않거나 통신하지 않음 - WKB18920

Article number: 18920To English version

증상

  • 통신이 없음

환경

  • 30cm 컬럼 히터/쿨러
  • ARC 시스템
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class

원인

Sample Manager 초기화시 컬럼 히터가 꺼졌습니다.

수정 또는 해결 방법

컬럼 히터/쿨러를 Sample Manager보다 먼저 켭니다.

추가 정보

id18920, column heater/cooler, UPAQSM, UPARBCH30, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBINARY, UPCH30, UPCH30A, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPDAARCB, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMPMGR

id18920, column heater/cooler, UPAQSM, UPARBCH30, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBINARY, UPCH30, UPCH30A, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPDAARCB, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMPMGR