Empower 프로젝트를 NuGenesis SDMS에 아카이브할 때 "The Project_Integrity.txt file wasn't backed up correctly" 오류가 표시됨 - WKB84942
Article number: 84942To English version
증상
- NNuGenesis Archive Agent 작업 로그 파일에 다음과 같은 메시지가 표시됩니다.
- Message........: DATE TIME: Archive scan with no files archived for command id 1. {'SDMSSERVER','Empower File Capture','Empower Archive'} {'Waters Corp.','Empower-3 Projects','0'}.{Summary : (Total Scanned Files = 0)(Start Time = DATE TIME)(End Time = DATE TIME)(Total Time = 0h 0m 1s)}.
'E ','DATE TIME ','0 ','0 ','The Project_Integrity.txt file wasn't backed up correctly. Please handle the project manually.EMPOWERSERVER.PARENTPROJECT\PROJECTNAME ','Waters.SDMS.WindowsService ','clsEmp2PreProc ','PerformProjBackup ','1 ',''
- Message........: DATE TIME: Archive scan with no files archived for command id 1. {'SDMSSERVER','Empower File Capture','Empower Archive'} {'Waters Corp.','Empower-3 Projects','0'}.{Summary : (Total Scanned Files = 0)(Start Time = DATE TIME)(End Time = DATE TIME)(Total Time = 0h 0m 1s)}.
- 로그에 명명된 프로젝트는 SDMS에 아카이브되지 않습니다.
- Empower는 명명된 프로젝트를 오류 없이 수동으로 백업할 수 있습니다.
- 로그에 명명된 프로젝트는 무결성 오류가 없습니다.
환경
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters R1 ~ R3
- Empower 3
- Empower 2
원인
NuGenesis RPC Service 계정 사용자(Windows)는 시스템이 백업을 시도하는 Empower 프로젝트의 Empower 원본 데이터 공유에 대해 최소한의 읽기 권한을 가지고 있지 않습니다.
수정 또는 해결 방법
NuGenesis File Capture 서버에 설정된 NuGenesis RPC Service 계정 사용자는 Empower 원본 데이터 공유에 대해 최소한의 읽기 권한이 있어야 합니다.
NuGenesis SDMS Archive에 설정된 모든 Empower 시스템에서 공유 권한을 적절하게 업데이트합니다.
추가 정보
id84942, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG