Empower 프로젝트를 NuGenesis SDMS에 아카이빙할 때 "Wrong number of channels was backed up from Empower database" 오류가 표시됨 - WKB59078
Article number: 59078To English version
증상
- NNuGenesis Archive Agent 작업 로그 파일에 다음과 같은 메시지가 표시됩니다.
- Message........: DATE TIME: Archive scan with no files archived for command id 1. {'SDMSSERVER','Empower File Capture','Empower Archive'} {'Waters Corp.','Empower-3 Projects','0'}.{Summary : (Total Scanned Files = 0)(Start Time = DATE TIME)(End Time = DATE TIME)(Total Time = 0h 0m 1s)}.
'E ','DATE TIME ','0 ','0 ','Wrong number of channels was backed up from Empower database. Please handle the project manually.EMPOWERSERVER.PARENTPROJECT\PROJECTNAME ','Waters.SDMS.WindowsService ','clsEmp2PreProc ','PerformProjBackup ','1 ',''
- Message........: DATE TIME: Archive scan with no files archived for command id 1. {'SDMSSERVER','Empower File Capture','Empower Archive'} {'Waters Corp.','Empower-3 Projects','0'}.{Summary : (Total Scanned Files = 0)(Start Time = DATE TIME)(End Time = DATE TIME)(Total Time = 0h 0m 1s)}.
- 로그에 명명된 프로젝트는 SDMS에 아카이브되지 않습니다.
- Empower는 명명된 프로젝트를 오류 없이 수동으로 백업할 수 있습니다.
- 로그에 명명된 프로젝트는 무결성 오류가 없습니다.
환경
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters R1 ~ R3
- Empower 3
- Empower 2
원인
원본 데이터 공유에 저장된 프로젝트의 DAT 파일 수가 프로젝트의 채널 수와 일치하지 않습니다. NuGenesis는 프로젝트에서 이 확인을 수행하여 Empower로부터 전체 백업을 받습니다.
수정 또는 해결 방법
- 프로젝트의 원본 데이터 공유에서 DAT 파일 수를 확인하고 그 수를 프로젝트의 채널 수와 비교합니다. SDMS가 프로젝트를 아카이빙하려면 번호가 일치해야 합니다.
- 숫자가 일치하지 않으면 SDMS가 프로젝트를 아카이빙할 수 없습니다.
추가 정보
id59078, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG