NuGenesis Archive Agent 로그 파일에 "Login to Millennium service failed" 오류가 표시됨 - WKB59019
증상
- Empower 데이터가 NuGenesis에 아카이빙되지 않습니다. 8 SR-2
환경
- NuGenesis 8 SDMS
- NuGenesis 8 Data Adapters Release 3 HF2
- Empower 3
원인
Empower 데이터베이스가 다운되었습니다.
수정 또는 해결 방법
Empower 데이터베이스를 시작합니다.
추가 정보
Archive Agent의 로그 파일인 "Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log"에 다음 메시지가 나타납니다.
Message........: 2018-11-21 10:36:34: Beginning execution of archive request 1.
Message........: 2018-11-21 10:36:34: Server: NUG, Project: Empower, Template: EmpClient.
"E ","2018/11/21 10:36:49 AM ","-1 ","-1 ","Login to Millennium service failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","ValidateInputs ","1 ",""
"E ","2018/11/21 10:36:49 AM ","-6 ","-6 ","Input Validation failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","2018/11/21 10:37:04 AM ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","2018/11/21 10:37:04 AM ","-1 ","-1 ","PreProcessor call failed for the remote server: WB-EMP-CL ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-11-21 10:37:04: EMC: PerformPreProcess error.
Message........: 2018-11-21 10:37:04: Archive scan with no files archived for command id 1. {"NUG","Empower","EmpClient"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 11/21/2018 10:36:34 AM)(End Time = 11/21/2018 10:37:04 AM)(Total Time = 0h 0m 30s)}.
Message........: 2018-11-21 10:37:04: Next scheduled runtime for archive request 1 is 2018-11-21 10:40:42.
이 오류 메시지의 또 다른 가능한 원인은 Error "Login to Millennium service failed" appears in the NuGenesis SDMS Archive Agent log(NuGenesis SDMS Archive Agent 로그에 "Login to Millennium service failed" 오류가 표시됨) - WKB26457에 나와 있습니다.
id59019, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG, 밸리데이션, 프로젝트