NuGenesis SDMS Archive Agent 오류 메시지 "Rollback: Performing rollback for . . ." - WKB49103
증상
- NuGenesis가 파일을 캡처하지 않습니다.
- Archive Agent 로그 파일에서 다음 오류 메시지가 표시됩니다.
Message........: DATE TIME: ExecuteMacroCommand: Could not restore the pending file (E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\70BCEC7E430FDD604B05978D9C591FB8). This file will not be archived.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\5F5750854EC851FA7A31BA8A43A7678B.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\54E70D504ABFC99846D607B17EC40000.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\AD365FA946DEE923E707B99A24481319.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\70BCEC7E430FDD604B05978D9C591FB8.
Message........: DATE TIME: ExecuteTemplateCommand: The group is rolled back because of errors.
환경
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
원인
AA에 의해 캡처되는 그룹의 모든 파일이 TE_TEMP에 복사되기 전에 시스템의 하드 드라이브에 사용 가능한 모든 공간이 사용되었습니다.
수정 또는 해결 방법
아직 해결 방법이 없습니다. File Capture 시스템은 후속 실행에서 기기 데이터 캡처를 시도합니다.
추가 정보
FC 서버 공간 부족이 자주 발생할 경우, TE_TEMP 폴더가 설정된 하드 드라이브의 크기를 늘려야 할 수 있습니다.