용매가 여전히 흐르는 상태에서 Empower의 실행이 중단됨 - WKB80771
Article number: 80771To English version
증상
- 실행 도중 실행이 중지됩니다.
- 샘플 세트가 시작되지 않습니다.
- Message Center에서 "System Suitability Fault"를 보고합니다.
환경
- Alliance HPLC 시스템
- 2695 Separations Module
- 2695D Dissolution Separations Module
- 2795 Separations Module
- ACQUITY APC
- ACQUITY APC Isocratic Solvent Manager
- ACQUITY APC Sample Manager - FTN
- ACQUITY Arc
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC
- ACQUITY UPLC BSM(예: 필요에 따라 다른 모듈 사용)
- ACQUITY UPLC Column Manager
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class Quaternary Solvent Manager(QSM)
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- PATROL UPLC 프로세스 분석 시스템
- ACQUITY UPLC Process Sample Manager
- ACQUITY UPLC 컬럼 히터
- ACQUITY UPLC PDA 검출기
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC Sample Organizer
- ACQUITY UPLC Tunable UV 검출기
- ACQUITY UPLC FLR 검출기
- ACQUITY UPLC RI 검출기
- Breeze 2
- Empower 3
- Empower 2
- Empower 3 퍼스널
- Empower 3 워크그룹
- Empower 3 엔터프라이즈
원인
Run Samples 페이지가 "Run and Process" 및 "Stop on Fault"로 설정되었습니다.
이 설정은 Empower가 Processing Method와 비교하여 주입을 확인하도록 합니다. 사양을 벗어나면 실행이 중지됩니다.
수정 또는 해결 방법
Run Samples 페이지를 "Run Only" 또는 "Continue on Fault"로 설정합니다.
추가 정보
id80771, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2790XE, 2790XEUP, 2795E, 2795UP, 2795XE, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, BREEZESW, column heater, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, SUPBZ, UPBINARY, 시스템적합성