답변

PFAS 분석용 Oasis WAX는 로트당 32 PFAS에 대해 Blank 테스트를 거쳤습니다. 테스트한 PFAS 및 허용 기준 농도 표는 추가 정보 섹션 에 첨부되어 있습니다.

추가 정보

 

CAS #

분석 물질

수락 한계

단위

375-22-4

Perfluorobutanoic acid (PFBA)

1

ng/L

2706-90-3

Perfluoropentanoic acid (PFPeA)

1

ng/L

307-24-4

Perfluorohexanoic acid  (PFHxA)

1

ng/L

375-85-9

Perfluoroheptanoic acid (PFHpA)

1

ng/L

335-67-1

Perfluorooctanoic acid (PFOA)

1

ng/L

375-95-1

Perfluorononanoic acid (PFNA)

1

ng/L

335-76-2

Perfluorodecanoic acid (PFDA)

1

ng/L

2058-94-8

PFUnA(Perfluoroundecanoic acid)

1

ng/L

307-55-1

PFDoA(Perfluorododecanoic acid)

1

ng/L

72629-94-8

Perfluorotridecanoic acid (PFTrA)

1

ng/L

376-06-7

Perfluorotetradecanoic acid (PFTDA)

1

ng/L

67905-19-5

Perfluorohexadecanoic acid (PFHxDA)

1

ng/L

16517-11-6

Perfluorooctadecanoic acid (PFODA)

1

ng/L

375-73-5

Perfluorobutanesulfonic acid (PFBS)

1

ng/L

2706-91-4

Perfluoropentanesulfonic acid (PFPeS)

1

ng/L

355-46-4

Perfluorohexanesulfonic acid (PFHxS)

1

ng/L

375-92-8

Perfluoroheptanesulfonic acid (PFHpS)

1

ng/L

1763-23-1

Perfluorooctanesulfonic acid  (PFOS)

1

ng/L

68259-12-1

Perfluorononanesulfonic acid (PFNS)

1

ng/L

335-77-3

Perfluorodecanesulfonic acid (PFDS)

1

ng/L

79780-39-5

Perfluorododecanesulfonic acid(PFDoS)

1

ng/L

754-91-6

Perfluorooctanesulfonamide (FOSA)

1

ng/L

2355-31-9

N-methylperfluorooctanesulfonamidoacetic acid (MeFOSAA)

10

ng/L

2991-50-6

N-ethylperfluorooctanesulfonamidoacetic acid (EtFOSAA)

10

ng/L

757124-72-4

1H,1H,2H,2H-perfluorohexanesulfonic acid(4:2 FTS)

10

ng/L

27619-97-2

1H,1H,2H,2H-perfluorooctanesulfonic acid(6:2 FTS)

10

ng/L

39108-34-4

1H,1H,2H,2H-perfluorodecanesulfonic acid (8:2 FTS)

10

ng/L

120226-60-0

1H,1H,2H,2H-perfluorododecanesulfonic acid (10:2 FTS)

1

ng/L

13252-13-6

Perfluoro(2-propoxypropanoic) acid (HFPO-DA 또는 Gen-X)

2

ng/L

73606-19-6

9-Chlorohexadecafluoro-3-oxanonane-1-sulfonate  (F-53B Major)

1

ng/L

83329-89-9

11-Chloroeicosafluoro-3-oxaundecane-1-sulfonate  (F-53B Minor)

1

ng/L

919005-14-4

ADONA

1